LCMS and LCMSMS Service: Properly Maintaining your Mass Spectrometer

Ted Palashis | March 4, 2020

Liquid chromatography-mass spectrometry, also referred to as LCMS, is an analytical chemistry technique that combines the mass analysis capabilities of mass spectrometry with the physical separation ability of liquid chromatography. Such a highly sensitive instrument needs to be serviced and maintained appropriately and effectively. Avoiding contamination can help prevent signal suppression, reduce background noise, minimize adduct formation, and increase its ability to identify spectrum complexity. In order to achieve the most accurate results, contamination prevention is crucial. To maximize your LCMS instrument sensitivity, we suggest the following:


Only Use Volatile Buffers

While some instrument manufacturers claim that involatile buffers like phosphate can be used, you should only use volatile buffers. The sampling orifice can become blocked and the voltages required by the electrostatic element can change. The conditions created by involatile buffers hamper optimum ion generation.


Check the Rotary Pump Oil

Rotary pumps consume minuscule amounts of oil as a part of their normal operation, and they must be topped off regularly. The rotary pumps can be exposed to a variety of different materials and solvents that can increase wear on the pump’s components. We advise draining and replacing the oil at the manufacturer’s recommended interval. Additionally, some models of these pumps can be ballasted to remove more volatile contaminants. If your pump has this feature, use the valve on the pump to clear it on a weekly basis to maintain the quality of the vacuum within the mass spectrometer.


Limit Your Flow Rate

While electrospray can allow newer equipment to utilize a higher flow rate, it is still advisable to use a lower flow rate. With an increased flow rate, more material enters the mass spectrometer, which increases the rate of contaminant buildup.


Using a Divert Valve

Using a divert valve, you can switch portions of the mobile phase to waste before it reaches the mass spectrometer. With a properly optimized method, this allows you to only send your analytes through the mass spectrometer, further preventing buildup of involatile components within the instrument.


Limit Your Sample Concentration

Loading a high concentration of your analyte onto the column can increase the visibility of smaller components, however, it can also quickly contaminate your source. Using a divert valve can divert the concentrated material to waste, helping to prevent contamination, however the best way to mitigate the risk of compromising your source is to limit your sample concentration when possible. If you must have the concentrated analyte go into the source, we recommend limiting the number of injections.


Do Not Vent Too Frequently

When you vent a mass spectrometer you increase the risk of wear and other component damage. These instruments perform best when they are constantly running. The turbo pump which creates a high vacuum is especially susceptible to damage from venting. The rush of air that accompanies the vacuum when it is engaged puts an intense amount of stress and strain on the pump’s vanes, specifically in the bearings, which wears the turbo down at a higher rate.


Maintaining a clean laboratory from the inside out of each asset is essential to continuing to produce the accurate results you need. Overbrook is an experienced, skilled, and educated provider of laboratory equipment services, especially when it comes to LCMS instrumentation.



CONTACT US
By Ted Palashis June 19, 2026
Cambridge is one of the most active life sciences hubs in the world. Driven by Kendall Square, MIT, Harvard, and hundreds of biotech trailblazers, the market is constantly evolving. For organizations expanding or consolidating, executing seamless lab relocation services in Cambridge is critical to maintaining a competitive edge. However, moving a laboratory in this historic, high-density market presents unique hurdles that standard commercial movers simply cannot handle. Why are Lab Relocations Unique in Cambridge? Unlike suburban research parks, Cambridge laboratories operate under tight spatial and regulatory constraints. Navigating laboratory relocation services in the Kendall Square and Greater Boston area requires managing specific urban logistical challenges: High-Density Logistics: Densely populated buildings with limited loading dock access and strict freight elevator schedules. Complex Stakeholder Alignment: Relocations require coordinated efforts among laboratory personnel, facility managers, specialized vendors, property owners, and municipal contractors. Strict Compliance & Infrastructure: Meeting rigid Cambridge bio-safety guidelines while integrating with shared building infrastructure (such as centralized vacuum or RO/DI water systems). The Golden Rule: Research Cannot Stop During a Lab Move Most life science organizations must migrate while research programs remain fully active. Project deadlines do not pause, and maintaining scientific productivity throughout the transition is paramount. "Many organizations compare laboratory relocations to changing the wheels on a moving car," says Ted Palashis , Founder of Overbrook Support Services. "Research doesn’t stop just because the physical space is changing. The true metric of success is how effectively you maintain scientific continuity, protect sample integrity, and minimize operational downtime." Key Steps for Successful Laboratory Equipment Planning The most common relocation bottlenecks occur after the equipment arrives at the new facility. Misaligned utilities or unverified clearances can stall a project for weeks. To prevent these delays, a specialized lab move requires a meticulous three-step asset assessment: Comprehensive Asset Audits : Conducting physical lab asset audits allows you to document every instrument, its current operational state, and its specific transport requirements (e.g., climate control, shock sensitivity). Utility Verification: Confirming that the new facility’s electrical, gas, data, and ventilation outputs precisely match manufacturer specifications. Addressing these details early prevents arriving at the new facility only to find misaligned utilities or unverified clearances that can stall a project for weeks. Spatial Workflow Mapping: Validating physical clearances (doorways, hallways, elevator capacities) and planning equipment placement. Utilizing precise spatial workflow mapping ensures that the layout optimizes daily laboratory operations from day one. Evaluating Facility Readiness Before Equipment Installation A laboratory may look complete from a construction standpoint yet be entirely unprepared for actual scientific operations. Before a single instrument is installed, specialized lab relocation services must verify that the facility is functionally operational. Coordinating an integrated laboratory fit-out ensures that utility capacities, service access, and vendor timelines are perfectly aligned before equipment arrives. This includes testing backup power systems, verifying gas line purities, checking HVAC air exchange rates, and ensuring that environmental monitoring systems are fully online. Addressing these infrastructure requirements early prevents costly startup delays. Comprehensive Lab Relocation Services for Cambridge Biotech The success of a Cambridge laboratory move is entirely defined by the strategy built before moving day. Organizations that invest in early equipment planning experience significantly less downtime and a much faster return to scientific discovery. A specialized lab moving checklist includes: Pre-move asset tagging and validation OEM vendor coordination for specialized instrument decommissioning/re-commissioning Chain-of-custody transport for sensitive biological samples and chemical reagents Post-move calibration and validation support At Overbrook Support Services , we provide tailored lab relocation services in Cambridge and the Greater Boston area. Our team focuses on continuity, compliance, and operational success, ensuring your science moves forward without interruption. For deeper insights into successful lab relocations and complex transitions, you might find this biotechnology enterprise course webcast with Ted Palashis highly relevant, as it features Overbrook's founder discussing real-world strategies for avoiding costly relocation mistakes .
Laboratory professional reviewing quality management and compliance
By Ted Palashis June 16, 2026
Learn how a robust Laboratory Quality Management System (QMS) helps laboratories improve compliance, reduce audit stress, strengthen documentation control, and support long-term operational growth.
Core components of a lab QMS integrated into daily laboratory operation
By Ted Palashis February 27, 2026
Understanding the core components of a Laboratory Quality Management System helps labs maintain compliance, audit readiness, and operational consistency across documentation, training, and equipment.
Show More